"The Vintner's Vault has been a major asset to our winery's success. We relied on them to set up our entire winery's operations from the ground up. Their wine making expertise, coupled with the equipment knowledge of Ryan Horn and his staff's professional expertise on hand and "locally" here in Paso. The Vintner's Vault's dedication to quality products and services provides the caliber of resources that will continue making Paso Robles one of the greates wine regions in the country."

Home > Winery Lab Supplies > Test Kits > Cash Still

Cash Still

Click to Enlarge

Cash Still
LAB-004823
Cash Still

Price:
$1,172.25

Add to Cart


Cash Still
A Cash Still is used to determine the Volatile Acidity of wine.  Our set-up comes without the chemicals, however we do supply them.  Please take a look below to see what chemicals are needed.

 
Equipment Included:
- Boiler
- Condenser
- Heating Element
- Stand + Rod
- Drain "T"

- Band Clamp
- Band Clamp Holder
- Latex Drain Tube (10ft)
- Latex Tubing (10ft)
- Pinch Clamps (2)
- 3 Finger Clamp
Place approximately 500 ml purified water in the outer chamber of the Cash still. Start
cold water flow through the condenser. Rinse the inner chamber. Place exactly 10 ml of sample in the inner chamber. Add 2 drops of Antifoam B and rinse the funnel into the
inner chamber with a few ml of water. Place an Erlenmeyer flask under the condenser to collect the distillate.
Set the 3-way valve to vent the outer chamber and apply power to the heating element.
When steam begins to escape, close the 3-way valve. Collect 125 ml of distillate.
Open the 3-way valve to vent the outer chamber and turn off the heating element power.
Refill the outer chamber with water. Open the 2-way valve to remove the contents of the inner chamber. Rinse the inner chamber twice with water before introducing additional
samples.
Add 3 drops of phenolphthalein indicator and titrate the distillate with 0.1 N NaOH to a
durable pink color. Add 1 ml starch indicator, 5 ml sulfuric acid and titrate with 0.02 N
Iodine to a faint blue color.
CALCULATIONS
Volatile acidity (gm/100 ml Acetic acid) =(0.60x mL NaOH x N NaOH)–(0.60 x mL I
= 0.60x[(mL NaOH x N NaOH) - (mL I
If NaOH is 0.10N and Iodine is 0.02N = (mL NaOH x 0.06) - (mL I

2 x N I2)2 x N I2)]2 x 0.012)

Winery-Brewery-Distillation Equipment & Supplies

 


Customers Who Bought This Item Also Bought:

Squirt Handle Wash Bottle

Squirt Handle Wash Bottle
16 ounce plastic bottle with graduations both in ounces and milliliters up to 12 oz/350ml

Price:
$4.99

Product Details Add to Cart

Aspiration Vacuum Kit So2

Aspiration Vacuum Kit So2
Vacuum Aspiration Kit

Price:
$430.00

Product Details Add to Cart

Your Cart Contains:

0 Item: $0.00

Please fill out the form below to sign up for our newsletter.
Copyright © 2019 The Vintner Vault | Website Development and Hosting by SMS Masterminds